In the table below are the channels we are currently working on.
| ADC channel |
Drawer |
LED adc |
LED ratio |
preamp adc |
preamp ratio |
preliminary diagnosis |
| 0_1_5 |
NW8 |
705 |
2.74 |
10297 |
4.24 |
| 0_2_1 |
NW7 |
-49 |
-0.13 |
682 |
0.28 |
broken preamp |
| 0_5_3 |
NW5 |
748 |
3.43 |
10322 |
4.22 |
| 0_8_2 |
NW3 |
182 |
0.31 |
696 |
0.28 |
broken preamp |
| 0_11_6 |
NW1 |
11 |
0.03 |
10400 |
4.21 |
| 1_5_3 |
NW13 |
745 |
4.97 |
10305 |
4.17 |
| 1_10_6 |
NW10 |
28 |
4.67 |
9799 |
4.12 |
| 1_11_5 |
NW9 |
630 |
1.81 |
10894 |
4.2 |
| 4_1_6 |
SW9 |
587 |
4.89 |
9819 |
4.13 |
hv low |
| 4_1_5 |
SW9 |
565 |
4.56 |
9807 |
146.37 |
hv low |
| 4_1_4 |
SW9 |
682 |
4.7 |
9676 |
3.9 |
hv low |
| 4_4_4 |
SW11 |
422 |
4.31 |
9678 |
156.1 |
| 4_10_1 |
SW15 |
728 |
4.89 |
9996 |
4.19 |
| 5_1_5 |
SW1 |
2156 |
5.03 |
255 |
0.11 |
loose pulser controller cable |
| 5_1_3 |
SW1 |
1379 |
4.79 |
263 |
0.11 |
loose pulser controller cable |
| 5_1_1 |
SW1 |
2017 |
4.8 |
0 |
0 |
loose pulser controller cable |
| 10_4_6 |
NE11 |
434 |
5.64 |
9395 |
4.05 |
| 11_1_5 |
NE1 |
56 |
0.08 |
10675 |
427 |
| 11_2_5 |
NE2 |
121 |
0.17 |
10533 |
87.78 |
The charts on this page represent our (relatively) current understanding of broken channels in the ICD. Channels can fail in 4 ways:
- The channel can have a mean ADC count < 750 in an LED run.
- The channel can have a mean ADC count < 750 in a preamp pulser run.
- The ratio of mean ADC in the LED run to reference LED run 216345 is less than 3.5.
- The ratio of mean ADC in the preamp pulser run to reference preamp pulser run 216711 is less than 3.5.
Each plot is labeled using run numbers in the form LED_preamppulser. The z-axis (color) is z=(cut-value)/cut. Thus a channel reporting 500 ADC counts would have a z value of .33 and a channel reporting a ratio of 3.0 would have a z value of .14. The closer to 0, the closer to passing the cut.
Since it can be difficult to see the x-axis labels, which are ADC crate_ADC card_Drawer ch., I have included a link to the .root file so you can zoom in as you like (click on the image).
*There is no preamp pulser data in this plot because the pulsers have broken.
jtorborg
Last modified: Tue May 16 09:44:24 CDT 2006